HW-Time-Slice Firmware-Filter Software vx2740 Steering Module Reports MVM MVM Vexos MVM-Bug listing MVM TRIUMF Local DS Prototype DS Cryogenic For Shifters BCIT-31 ChronoBox Run Operation DS-DAQ
  CERN DS-Proto0 read-only backup, Page 2 of 8  Not logged in ELOG logo
ID Date Authorup Type Category Subject Text Attachments
  69   04 Nov 2019 09:02 Alex/Yi/BenRoutineGeneralrun 865, fields onGas pocket: ON

Fields: wire
gate 3780 V, 1st ring: 4500 V, cathode: 7000
  
  49   28 Oct 2019 08:33 Ben SmithProblem FixedSoftwareNeed python3 package tkinter on DAQ pcI have installed the python3-tkinter package
from yum. I checked that "import tkinter"
works from a python3 prompt.
  
  50   28 Oct 2019 14:10 Ben SmithProblem FixedSoftwareConverters installed in VME crate> V1725 board #0 logic level is set to TTL
(boards #1, #2, #3 to NIM) 
  
  53   01 Nov 2019 06:07 Ben SmithProblem FixedSoftwareRestoration of network settingsFor a few days we were unable to access ds-proto-daq
remotely. We were also unable to reach the
outside world from ds-proto-daq.
  
  54   01 Nov 2019 06:15 Ben SmithProblem FixedSoftwareRestoration of elogAfter booting, the elog was only showing logs
from before July, and was missing the "For
Shifters" category. It appears we were loading
  
  66   04 Nov 2019 07:57 Ben SmithConfigurationSoftwarePackage installationsThis morning I installed the xrootd-client
and python3-devel packages from yum. For
the latter to work, I needed to do a yum
  
  67   04 Nov 2019 08:00 Ben SmithConfigurationMIDASChange of location of history files/home was getting very full on ds-proto-daq,
so I moved the history files onto the /data
disk. There is now a symlink from ~/online/history
  
  68   04 Nov 2019 08:03 Ben SmithRoutineSoftwareNew script for setting self-trigger thresholdsI have created a new script that simplifies
the adjustment of self-trigger thresholds
(e.g. to change between "noise" runs with
  
  70   05 Nov 2019 01:22 Ben SmithConfigurationDigitizerV1725 board configThe V1725s have been changed to have board
config 0x50 rather than 0x10. This means
that they will now trigger on the leading
  
  74   05 Nov 2019 05:43 Ben SmithRoutineSoftwareNew script to equalize baselinesI've written a new script that will set all
the V1725 baselines to the same level, by
adjusting the DAC values. All the baselines
  
  87   06 Nov 2019 05:58 Ben SmithConfigurationHardwareCAEN VME Crate DHCPThe CAEN VME crate now gets its IP address
from the ds-proto-daq DHCP server. To do
this, I set the IP address of the crate to
  
  89   06 Nov 2019 12:28 Ben SmithProblem FixedGeneralEvening data taking, purity check> Not sure if the trigger threshold is correctly
set, there is an error:
  
  95   08 Nov 2019 02:21 Ben SmithProblem FixedSoftwareIncreased max waveform length that we can takeI've increased various thresholds so that
we can now take 3000us (3ms) waveforms.
  
  96   08 Nov 2019 02:31 Ben SmithRoutineSoftwareNew "run type" selection pageI made a new page that makes it easier to
switch between laser/noise/physics runs.
  
  97   08 Nov 2019 02:48 Ben SmithConfigurationMIDASEnabled dotfile naming in LoggerI have changed `/Logger/Channels/0/Settings/Filename`
to start with a dot, so files will be hidden
until they are complete. 
  
  98   08 Nov 2019 05:14 Ben SmithConfigurationHardwareAdding darkside01 to the networkThe analysis server from Roma (darkside01)
is now connected to the local network. It
is not yet connected to CERN.
  
  99   08 Nov 2019 08:35 Ben SmithConfigurationHardwareFailed to talk to steering module from DAQ PCI have failed to communicate with the steering
module arduino from the DAQ PC.
  
  100   08 Nov 2019 09:09 Ben SmithConfigurationHardwareAdding darkside01 to the networkThe cable has been fixed. Socket 0001/007
is still not active.
  
  120   12 Nov 2019 00:23 Ben SmithProblemDigitizerV1725 error, runs keep crashingIt looks like there's a memory leak in the
high voltage driver, and we were running
out of memory (using a lot of swap). That
  
  126   13 Nov 2019 08:15 Ben SmithRoutineSoftwareDon't worry about "Param Not Found Type" error messagesWhen we start the HV program (CAEN_SY4527),
it spews error messages of the form:
"16:58:30.035 2019/11/13 [CAEN_SY4527,ERROR]
  
ELOG V3.1.4-cb3afcd8