  128   14 Nov 2019 00:09 Ben SmithProblemHardwareChronobox not serving human webpage
The chronobox isn't serving the human-interactive webpage that should be visible at https://m-darkside.web.cern.ch/chronobox/ - it responds with "File not found". Oddly, it is still responding to API calls at /read_var, /write_var etc.

So shifters can still change between laser/noise/physics runs using the "Run type" page at https://m-darkside.web.cern.ch/?cmd=custom&page=Run%20type (which uses the API), but can't as easily monitor / sanity check the chronobox behaviour.

In the first instance, I think it would be useful to power-cycle the chronobox. If that doesn't work, I'm not sure how to proceed.
  133   14 Nov 2019 08:14 Ben SmithRoutineGeneralAutomatic copying to darkside01
I have set up automatic copying of data from ds-proto-daq to darkside01, using the lazy logger.

There are 2 lazy logger channels set up - one for lz4 files and one for gz files. It will take some time for all the files to copied over. Note that lazy logger writes a message into the midas log for every file that it copies. The lazy logger channels can be started from the Program page like all the other midas-related programs.

Once files have been copied to darkside01 (by lazy logger) and to EOS (by Pablo/Edgar) we can delete files from ds-proto-daq. We will probably not be able to do this until tomorrow (copying is quite slow).
  143   18 Nov 2019 07:20 Ben SmithRoutineGeneralDisk cleanup
Lazy logger has caught up with the backlog, and continues to transfer new files to darkside01 as they are written.
EOS currently contains data for runs 822-1201.

I have deleted the following runs from ds-proto-daq:
* runs < 700 that were taken with gzip compression
* runs 800-1100 that are available on EOS and darkside01

The following files remain:
* runs 700-800 that haven't been copied to EOS (but are on darkside01). If they're of interest, they should be copied to EOS.
* runs 1200+ that haven't been copied to EOS yet (but will be soon).

Disk usage has improved from 85% to 33%.
  149   16 Mar 2020 14:21 Ben SmithConfigurationSoftwareNew CAEN HV frontend
A few months ago I created a new midas frontend for CAEN HV crates. It behaves more like what users expect - changes made through other interfaces (e.g. ssh or the Java-based GUI) will be reflected in the ODB. The ODB structure is created dynamically based on which modules are present in the crate, and the parameters that each module supports (e.g. whether you specify a ramp-up rate in V/s or a ramp-up time in s).

When I first wrote the frontend, ds-proto-daq was offline. Pierre reminded me today that I should actually deploy the new code.

The new program code is at ~/online/dsproto_sy4527/caen_hv*. The executable is at ~/online/bin/caenhv_fe. The frontend is available on the "Programs" page in mhttpd. If we want to run the old HV frontend, it is still available at ~/online/bin/sy4527.

Connection parameters for the frontend are found in the ODB at /Equipment/CAEN_HV/Settings/Global. Currently the crate appears to be off, so the frontend will refuse to start properly as it cannot connect to 192.168.1.9. If the frontend is started with the crate on, it will create more directories in /Equipment/CAEN_HV/Settings (one for each module). The "Voltage" page on the experiment now points to a generic webpage that will work with arbitrary combinations of modules.
  150   08 Sep 2020 12:46 Ben SmithConfigurationOtherNew backup ELOG location
I have added a backup of the CERN proto-0 ELOG at TRIUMF, in case ds-proto-daq is unavailable in future.

The backup ELOG is at https://ladd00.triumf.ca/elog-ds/. There are other logbooks hosted in the same place (requiring a username/password), but the CERN backup logbooks (at the bottom of the list) are available to all. I have configured things so that the TRIUMF replicas are read-only. To add new entries one must still use the CERN logbooks.

The logbooks are rsynced every hour via cron.
  19   06 Mar 2019 18:19 Bryerton ShawRoutineTriggerSetup of chronobox
The SDcard is currently required for operation of the device, the ext3/4 filesystem will immediately fail upon removal.


> Summary of setup of chronobox on network (mostly done by Pierre):
> 
> 1) Hook up USB connection from chronobox to ds-proto-daq.  Start serial-USB connection by doing
> 
> /home/dsproto/online/dsproto_daq/serialusb
> 
> can login through serial link.  Added the ssh key for dsproto@ds-proto-daq to the authorized_keys file on
> chronobox linux.
> 
> ctrl-a, ctrl-x to stop picocom
> 
> 2) On ds-proto-daq, setup dhcpd server.  Configuration in this file
> 
> /etc/dhcp/dhcpd.conf
> 
> dhpcd is bond to the second NIC only.  Configure DHCP to give IP 192.168.1.3 to chronobox.  In /etc/hosts set
> 192.168.1.3 to have hostname chronobox
> 
> Power cycle chronobox; it successfully gets IP:
> 
> [dsproto@ds-proto-daq dsproto_daq]$ ping chronobox
> PING chronobox (192.168.1.3) 56(84) bytes of data.
> 64 bytes from chronobox (192.168.1.3): icmp_seq=1 ttl=64 time=0.180 ms
> 
> 3) Can see the chronobox webpage locally as
> 
> http://chronobox:8080
> 
> 4) Some esper tool to access data:  did following to setup esper
> 
>   734  yum install python2-pip
>   735  pip install esper-tool
>   736  yum install ncurses
>   738  yum install ncurses-devel
>   739  pip install esper-tool
> 
> Can then connect to the chronobox by doing
> 
> esper-tool interactive http://chronobox:8080
> 
> This is as much as I understand at this point... more exploring now...
> 
> 5) Pierre unplugged the SD card to take it back to Bryerton's room.  But I guess this was bad.
> 
> lots of errors on serialUSB link now and the webpage doesn't work anymore:
> 
> [ 1246.665269] blk_partition_remap: fail for partition 2
> Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.904770] EXT4-fs error: 339 callbacks suppressed
> [ 1246.679162] blk_partition_remap: fail for partition 2
> Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.904780] EXT4-fs error (device mmcblk0p2):
> ext4_find_entry:1437: inode #2: com[ 1246.692884] blk_partition_remap: fail for partition 2
> m syslogd: reading directory lblock 0
> [ 1246.709020] blk_partition_remap: fail for partition 2
> Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.929199] blk_partition_remap: fail for partition 2
> [ 1246.717394] blk_partition_remap: fail for partition 2
> Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.942971] EXT4-fs error (device mmcblk0p2):
> ext4_find_entry:1437: inode #2: com[ 1246.731255] blk_partition_remap: fail for partition 2
> m syslogd: reading directory lblock 0
> Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.953760] blk_partition_remap: fail for [ 1246.747337]
> blk_partition_remap: fail for partition 2
> partition 2
> Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.967532] EXT4-fs error (device mmcblk0p2):
> ext4_find_entry:1437: [ 1246.763481] blk_partition_remap: fail for partition 2
> inode #2: comm syslogd: reading directory lblock 0
> Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.978320] blk_partition_rem[ 1246.779511] blk_partition_remap:
> fail for partition 2
> ap: fail for partition 2
> Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.992091] EXT4-fs error (device mmcblk0p2): ext4_find[
> 1246.795586] blk_partition_remap: fail for partition 2
> _entry:1437: inode #2: comm syslogd: reading directory lblock 0
> Jan  1 00:20:46 buildroot user.warn kernel: [ 1243.002885] blk_[ 1246.811671] blk_partition_remap: fail for
> partition 2
> partition_remap: fail for partition 2
> Jan  1 00:20:46 buildroot user.crit kernel: [ 1243.016656] EXT4-fs error (device mmcblk0p[ 1246.827750]
> blk_partition_remap: fail for partition 2
  112   10 Nov 2019 12:15 E. SanchezProblemMIDASProblem DAQ

Error during data taking  1573416708 21:11:48.151 2019/11/10 [feov1725MTI00,INFO] V1725 PLL loss lock Board:0 (vmeAcq=0x0)

It stop saving event but run in process. Impossible to stop the run. I tried to reset feov1725MTl00. After that error: 

1573416857 21:14:17.792 2019/11/10 [feChronoEsper,INFO] Client 'feov1725MTI00' on database 'ODB' removed by db_cleanup called by cm_periodic_tasks because pid 9724 does not exist

Run stopped after some time. Error 21:21:19.121 2019/11/10 [mhttpd,ERROR] [midas.cxx:3951:cm_transition_call,ERROR] cannot connect to client "feov1725MTI00" on host localhost, port 40944, status 503   ╳

Trying to restart feov1725MTI00. Status initializing

It is not possible to initialize 

 

 

  129   14 Nov 2019 00:33 E. SanchezRoutineGeneralMorning data taking

It is not possible to access to the Chronobox page

Run 1109 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 1110 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Run 1112

Gas pocket: ON (thinkness 7mm)

Fields: OFF 

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1113

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Run 1114

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Run 1115

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

 

 

 

  130   14 Nov 2019 03:51 E. SanchezRoutineGeneralS1 data taking without field

Run 1133

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1134

Gas pocket: OFF

Fields: OFF

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1135

Gas pocket: OFF

Fields: OFF

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1136

Gas pocket: OFF

Fields: OFF

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1137

Gas pocket: OFF

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1138

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1139

Gas pocket: OFF

Fields: OFF

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

 

Run 1140

Gas pocket: OFF

Fields: OFF

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1141

Gas pocket: OFF

Fields: OFF

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1142

Gas pocket: OFF

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1143

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1144

Gas pocket: OFF

Fields: OFF

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1145

Gas pocket: OFF

Fields: OFF

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1146

Gas pocket: OFF

Fields: OFF

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1147

Gas pocket: OFF

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

  132   14 Nov 2019 07:42 E. SanchezRoutineGeneralS1 data taking with field

Run 1148

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

problem DAQ, forced to stop manually

Run 1149

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 380 Hz

Run 1150

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 310 Hz

Run 1151

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  250 Hz

Run 1152

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  150 Hz

Run 1153

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  90 Hz

Run 1154

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  390 Hz

Run 1155

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

Run 1156

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

Run 1157

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:   400 Hz

Run 1158

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:   400 Hz

Run 1159

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1160

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1161

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1163

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Run 1164

Gas pocket: OFF

Fields: 500 V/cm

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

  135   15 Nov 2019 10:48 E. SanchezRoutineGeneralS1 data taking proposed by Marco

Run 1171

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1172 Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1173

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 4 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1174

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend: 5x16 = 80ns TPC

Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1175

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 6 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1176

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 500 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 2 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1177

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 500 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 3 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1178

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 500 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 4 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1179 Gas pocket: OFF

Fields: 0 V/cm

Threshold: 500 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1180

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 500 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Coincidence: 6 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 20k

Run 1181

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 200us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 100k

Problem with DAQ, forced to stop

Run 1182

Gas pocket: OFF

Fields: 0 V/cm

Threshold: 200 ADCc below baseline

Trace length: 200us total, 5 us pre-trigger

Coincidence: 5 PDMs

Threshold extend: 5x16 = 80ns

TPC Pressure: 80 mbarg

SiPM HV: 65 V

Number of events: 100k

  101   08 Nov 2019 10:07 Edgar SanchezRoutineGeneralMonitoring Data Taking

Run 995

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  230 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

Comment: TPC pressure increased up to 290 mbarg

Run 996

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  225 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

Comment: TPC pressure increased up to 240 mbarg

Run 997

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

Run 998

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

  108   10 Nov 2019 08:04 Edgar SanchezRoutineGeneralS1 data taking with field

Run 1032

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 800 Hz

Run 1033

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 350 Hz

Run 1034

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 250 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 260 Hz

Run 1035

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 200 Hz

Run 1036

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 130 Hz

 

"ERROR INFO: 17:19:23.176 2019/11/10 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."

Run 1037

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 800 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  110 Hz

Run 1038

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  80 Hz

Run 1039

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 150 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

 

  109   10 Nov 2019 08:40 Edgar SanchezRoutineGeneralS1 data taking without field

Run 1041

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 360 Hz

Run 1042

Gas pocket: OFF

Fields: OFF

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 340 Hz

Run 1043

Gas pocket: OFF

Fields: OFF

Threshold: 250 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 290 Hz

Run 1044

Gas pocket: OFF

Fields: OFF

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 240 Hz

Run 1045

Gas pocket: OFF

Fields: OFF

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 170 Hz

Run 1046

Gas pocket: OFF

Fields: OFF

Threshold: 800 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  140 Hz

Run 1047

Gas pocket: OFF

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  110 Hz

Run 1048

Gas pocket: OFF

Fields: OFF

Threshold: 150 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

Run 1049

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

  110   10 Nov 2019 11:18 Edgar SanchezRoutineGeneralAfternoon system monitoring

Run 1050

Gas pocket: ON (unknow)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  245 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1051

Gas pocket: ON (unknow)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1052

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  245 mbarg

SiPM HV: 65 V

Number of events: 10k

Stopped. Problem with DAQ

  111   10 Nov 2019 11:19 Edgar SanchezRoutineGeneralS1 + S2 short time window

Run 1054

Gas pocket: ON (unknow)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 100 ADC below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 400 Hz 

 

 

  122   12 Nov 2019 09:22 Edgar SanchezRoutineGeneralAfternoon system monitoring

Run 1090

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  220 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1091

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 400 ADCc below baseline

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  220 mbarg

SiPM HV: 65 V

Number of events: 10k

  123   12 Nov 2019 09:30 Edgar SanchezRoutineGeneralAfternoon system monitoring

Run 1092

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  220 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1093

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  220 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1094

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  220 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1095

Gas pocket: ON (thinkness 7mm)

Fields: OFF 

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  220 mbarg

SiPM HV: 65 V

Number of events: 10k

  124   12 Nov 2019 10:08 Edgar SanchezRoutineGeneralLaser run

Run 1096 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 1097 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

Problem: The access to the Chronobox page is not possible.  Rate 400 Hz 

  142   18 Nov 2019 07:01 Julie / EdgarRoutineGeneralRandom Trigger runs suggested by Simone, S1 data

PDM 15 and PDM 19 strange behaviour, lot of noise

Run 1239  Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1240  Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1241 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1242 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1243 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1244 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1245 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1246 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1247 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1248 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1249 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1250 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1251 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

Run 1252 Physics run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Threshold: external clock,  set to 50 Hz

Trace length: 200us, 100us pre-trigger

No field

SIPM HV: 65V

Number of events: 40k

 