Run 1032
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 800 Hz
Run 1033
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 350 Hz
Run 1034
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 250 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 260 Hz
Run 1035
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 400 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 200 Hz
Run 1036
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 600 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 130 Hz
"ERROR INFO: 17:19:23.176 2019/11/10 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."
Run 1037
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 800 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 110 Hz
Run 1038
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 80 Hz
Run 1039
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 150 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
|