  31   11 Jul 2019 15:52 ThomasRoutineTriggerinvert first chronobox busy signal
The DAQ seems to be ready for tests with proto-0 tomorrow.  

I had to invert the first busy input in order to get chronobox to produce triggers.  I modified the setup script

setup_chronobox.sh

so this is now the default.  Not sure why necessary.

Another interesting fact; it seems the chronobox only asks for a DHCP IP when it first boots.  I think that
chronobox and ds-proto-daq were rebooted at the same time; ds-proto-daq dhcp server was probably not running
when chronobox asked for IP.  chronobox got IP fine when it was power cycled.
  7   21 Jan 2019 13:05 Thomas, PierreRoutineDigitizerTests of V1725 baseline and triggering
A little progress on a couple fronts:

1) We got the baseline DAC setting working correctly from both the CAEN command line tool and from the MIDAS
frontend.  Needed to add some extra sleep before the DAC settings were modified.

2) We figured out how to set the threshold for self-triggering on each channel. It is actually register 0x1n60
for this firmware.  The threshold is defined relative to the online calculated baseline. The threshold is
settable from the ODB.  However, the polarity of the triggering is currently hard-coded to be negative (bit 6 of
register 0x1n64).

3) We could confirm that the LVDS outputs from the V1725 appeared and disappeared as expected when the
self-trigger threshold was changed.

4) We added some basic documentation in Markdown here:

https://bitbucket.org/ttriumfdaq/dsproto_daq/src/master/
  18   06 Mar 2019 15:10 Thomas, PierreRoutineTriggerSetup of chronobox
Summary of setup of chronobox on network (mostly done by Pierre):

1) Hook up USB connection from chronobox to ds-proto-daq.  Start serial-USB connection by doing

/home/dsproto/online/dsproto_daq/serialusb

can login through serial link.  Added the ssh key for dsproto@ds-proto-daq to the authorized_keys file on
chronobox linux.

ctrl-a, ctrl-x to stop picocom

2) On ds-proto-daq, setup dhcpd server.  Configuration in this file

/etc/dhcp/dhcpd.conf

dhpcd is bond to the second NIC only.  Configure DHCP to give IP 192.168.1.3 to chronobox.  In /etc/hosts set
192.168.1.3 to have hostname chronobox

Power cycle chronobox; it successfully gets IP:

[dsproto@ds-proto-daq dsproto_daq]$ ping chronobox
PING chronobox (192.168.1.3) 56(84) bytes of data.
64 bytes from chronobox (192.168.1.3): icmp_seq=1 ttl=64 time=0.180 ms

3) Can see the chronobox webpage locally as

http://chronobox:8080

4) Some esper tool to access data:  did following to setup esper

  734  yum install python2-pip
  735  pip install esper-tool
  736  yum install ncurses
  738  yum install ncurses-devel
  739  pip install esper-tool

Can then connect to the chronobox by doing

esper-tool interactive http://chronobox:8080

This is as much as I understand at this point... more exploring now...

5) Pierre unplugged the SD card to take it back to Bryerton's room.  But I guess this was bad.

lots of errors on serialUSB link now and the webpage doesn't work anymore:

[ 1246.665269] blk_partition_remap: fail for partition 2
Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.904770] EXT4-fs error: 339 callbacks suppressed
[ 1246.679162] blk_partition_remap: fail for partition 2
Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.904780] EXT4-fs error (device mmcblk0p2):
ext4_find_entry:1437: inode #2: com[ 1246.692884] blk_partition_remap: fail for partition 2
m syslogd: reading directory lblock 0
[ 1246.709020] blk_partition_remap: fail for partition 2
Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.929199] blk_partition_remap: fail for partition 2
[ 1246.717394] blk_partition_remap: fail for partition 2
Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.942971] EXT4-fs error (device mmcblk0p2):
ext4_find_entry:1437: inode #2: com[ 1246.731255] blk_partition_remap: fail for partition 2
m syslogd: reading directory lblock 0
Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.953760] blk_partition_remap: fail for [ 1246.747337]
blk_partition_remap: fail for partition 2
partition 2
Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.967532] EXT4-fs error (device mmcblk0p2):
ext4_find_entry:1437: [ 1246.763481] blk_partition_remap: fail for partition 2
inode #2: comm syslogd: reading directory lblock 0
Jan  1 00:20:46 buildroot user.warn kernel: [ 1242.978320] blk_partition_rem[ 1246.779511] blk_partition_remap:
fail for partition 2
ap: fail for partition 2
Jan  1 00:20:46 buildroot user.crit kernel: [ 1242.992091] EXT4-fs error (device mmcblk0p2): ext4_find[
1246.795586] blk_partition_remap: fail for partition 2
_entry:1437: inode #2: comm syslogd: reading directory lblock 0
Jan  1 00:20:46 buildroot user.warn kernel: [ 1243.002885] blk_[ 1246.811671] blk_partition_remap: fail for
partition 2
partition_remap: fail for partition 2
Jan  1 00:20:46 buildroot user.crit kernel: [ 1243.016656] EXT4-fs error (device mmcblk0p[ 1246.827750]
blk_partition_remap: fail for partition 2
  8   28 Jan 2019 01:44 ThomsRoutineSoftwareAdded web display of V1725 waveforms
I added javascript webdisplay of the V1725 waveforms to the Darkside setup.  You can see the waveform display here:

https://ds-proto-daq.triumf.ca/CS/webdisplay

You can also see the webdisplay that comes built into the JSROOT webserver here:

https://ds-proto-daq.triumf.ca/rootana/

For Pierre: I went through and reset the baselines for all the channels.
  151   14 Oct 2020 08:39 Tom ThorpeRoutineGeneralLog regarding MB2 removal from Proto-0

Drive folder with data files and some pictures:  https://drive.google.com/drive/folders/1Sr27kQTOO2kciBY7v-wYBP06tWLVEjiP

Lab notes @ CERN

11/10/2020

11:00 – First look at MB2.  Noticed on some of the PDMs that some of the tile PCB are more delaminated than others.  PDM 25 also had the FEB connectors more inserted than any other PDMs.  No other differences noticed.

11:30 – Connected SMU and switching matrix to MB2.  Power w/o steering module.  Checked that all channels match by running photo current tests with and w/o a flashlight on one corner.  Using the computer form Pisa with Labview scripts from Matteo.  Everything is ok.

12:00 – Replace plastic shields around Proto-0 for IV curve measurements in dark conditions.  PDM 1 breakdown is around 63V.  Looks ok.  PDM 2 is over 65V…. Seems high.

12:30 - PDM 6 People walking, disrupted the light conditions.  IV curve will be jagged.  Will go for lunch.  Closing door and turning off exterior lights.  Current will decrease

15:30 – IV curves look fine.  …/MB2_retriveval_CERN/data_files/iv_warm_1

16:00 – Took forward IV curves.  Saved as …/MB2_retriveval_CERN/data_files/forward_warm_1.  They look fine.  Checked old pictures of the lamination of the tile PCB and they have always varied in thickness, from before the LNGS LN test.

16:45 – Power FEBs on to look at noise spectra with 20V bias on the tiles.  

18:00 – PDM 2, 4, 6, and 25 have very odd power spectra.  George said they look like they’re not connected.

18:30 – Saving FFT to disk for analysis.  Saved as …/MB2_retriveval_CERN/data_files/noise_warm_1

19:00 – PDM 2,4,6 and 25 have one of the differential lines disconnected.  Using a light and looking at the differential outputs on the scope, this is obvious.  20V bias and FEBs powered at 5V.  Saved as …/MB2_retriveval_CERN/data_files/waveform_light_1

19:30 – Started another round of reverse IV curves in dark conditions.  Saved as …/MB2_retriveval_CERN/data_files/iv_warm_2

21:15 – IV still going.  EP will come back to shut off the LV supplies

 

12/12/2020

10:00 – Checking connectivity of all the differential lines.  One side of PDM 6 is not connected.  2, 4, and 25 are connected.  Doesn’t explain what we saw yesterday.  Channel 2 and 4 now are responding to light on both differential lines.  The only thing that has changed is unplugging the external signal cable and plugging it back in.  PDM 6 is the only one now not responding on both lines.

10:30 – Checking the power spectra.  PDM 4 and 6 are bad on both differential lines.  PDM 2 looks fine.  Seems to be non-repeatable.

11:00 – Disconnected the signal patch board from the finger strip.  Holding the patch and reconnecting the entire signal path and pinging now shows PDM 6 and 25 have one differential line disconnected.

11:30 – Checking the power spectra on every PDM channel directly from the fingerstrip with a flat cable.

12:30 – PDMs 2, 4, 6 power spectra from the fingerstrip look normal.  Both connections are there.  Light response is normal.  Channel 13, however, now has one differential line disconnected and the other looks abnormal.  Saving waveform data with 60Hz light on as waveform_light_2.

All of this is pointing to multiple connection issues…

15:00 – It looks like PDM 13 is disconnected from the fingerstrip, or at least it is pulled further away than the others.  This is likely also due to the removal of the patch.  PDM 2 also looks pulled a bit further.

15:30 – Unmounted MB2 and put on the stand to do an optical inspection.  The microscope didn’t allow a close enough look at the wirebonds to search for any broken ones.  A close look by eye didn’t reveal any obvious damage.

16:00 – Packed everything up.  MB2 needs a closer look, likely needs to be disassembled.

Attachment 1: forward_warm_1.png
forward_warm_1.png
Attachment 2: iv_warm_1.png
iv_warm_1.png
Attachment 3: iv_warm_2.png
iv_warm_2.png
  104   09 Nov 2019 00:13 Xiang XiaoRoutineGeneralMorning data taking

"ERROR INFO: 10:38:10.924 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."

Emailed to Ben.

 

Run 999

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  233 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1000

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  235 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1001

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  238 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1002

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  238 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1003 Laser run (manually stopped)

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Comment: Triger rate too low (~8 Hz), manually stopped

 

Run 1004 Laser run (manually stopped)

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Comment: Triger rate too low (~8 Hz), manually stopped

 

  127   13 Nov 2019 09:24 Xiang XiaoRoutineGeneralEvening data taking

Run 1105

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  265 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1106

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  252 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1107

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1108

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  242 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Comment: Webpage of Chronobox doens't work, skipped Laser runs.

  114   11 Nov 2019 00:47 Yi / EdgarProblem FixedDigitizerProblem fixed by restarting the VME crate
  115   11 Nov 2019 01:12 Yi / EdgarRoutineGeneralMorning data taking

Run 1065

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1066

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1067

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

  139   18 Nov 2019 00:18 Yi / EdgarRoutineGeneraldual phase data taking

We still have the noise issue.

Run 1219

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1220

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1221

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1222

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1223

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1224

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1226

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 40k

 

Run 1228

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 15k

 

Run 1229

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 40k

  71   05 Nov 2019 04:48 Yi WangRoutineGeneralRun 873

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar)

Fields: drift 500V/cm extraction 3.78kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment: try to trigger on S2

  72   05 Nov 2019 04:57 Yi WangRoutineGeneralRun 875

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment: compare to Run 873

  73   05 Nov 2019 05:09 Yi WangRoutineGeneralRun 877

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment: compare to Run 873, 875

  86   05 Nov 2019 23:58 Yi WangRoutineGeneralMorning data taking, purity check

Run 917

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 918

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 919

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 920

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

  88   06 Nov 2019 11:22 Yi WangRoutineGeneralEvening data taking, purity check

Not sure if the trigger threshold is correctly set, there is an error:

Traceback (most recent call last):
  File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 71, in <module>
    main()
  File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 47, in main
    (mean, rms) = fit_res
ValueError: too many values to unpack (expected 2)

the trigger threshold for all the runs in this Elog is 2699 ADCc below the baseline (see Elog 89)
 

Run 943

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 945

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 946

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 948

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 949

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 951

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 952

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 954

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

  94   07 Nov 2019 23:48 Yi WangRoutineGeneralmorning data taking

Run 973

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 974

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 975

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 976

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 978 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 979 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

After laser run:

Switch Laser OFF

In Chronobox:

DIsable "External Triger Enable [ext_trig_ena]0\[0]

Input "0x3F3F3F7F" in "Enable Channel [ch_enable] 

 

  107   10 Nov 2019 00:06 Yi WangRoutineGeneralmorning data taking

Run 1022

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  159 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1024

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  186 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1025

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  207 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1026

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  225 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1028 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 1029 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

  113   11 Nov 2019 00:37 Yi WangProblemMIDAScannot connect to rootana

The connection to rootana is lost.

No data taking can be started.

  118   11 Nov 2019 23:57 Yi WangProblem FixedTriggerID 117 problem is fixed

The problem reported in ID 117 is somehow fixed this morning.

  119   12 Nov 2019 00:16 Yi WangProblemDigitizerV1725 error, runs keep crashing

09:15:33.955 2019/11/12 [feov1725MTI00,ERROR] [feoV1725.cxx:663:link_thread,ERROR] Exiting thread 3 with error