ID |
Date |
Author |
Type |
Category |
Subject |
Text |
|
119
|
12 Nov 2019 00:16 |
Yi Wang | Problem | Digitizer | V1725 error, runs keep crashing | 09:15:33.955 2019/11/12 [feov1725MTI00,ERROR]
[feoV1725.cxx:663:link_thread,ERROR] Exiting
thread 3 with error |
|
120
|
12 Nov 2019 00:23 |
Ben Smith | Problem | Digitizer | V1725 error, runs keep crashing | It looks like there's a memory leak in the
high voltage driver, and we were running
out of memory (using a lot of swap). That |
|
128
|
14 Nov 2019 00:09 |
Ben Smith | Problem | Hardware | Chronobox not serving human webpage | The chronobox isn't serving the human-interactive
webpage that should be visible at https://m-darkside.web.cern.ch/chronobox/
- it responds with "File not found". Oddly, |
|
1
|
11 Dec 2018 15:17 |
Thomas | Configuration | Software | Setup elog for ds-proto-daq |
1) Install and tweak elog:
|
|
2
|
11 Dec 2018 15:20 |
Thomas | Configuration | Software | Setup elog for ds-proto-daq |
1) Install and tweak elog:
|
|
9
|
31 Jan 2019 15:18 |
Pierre | Configuration | Trigger | Trigger rate | Somehow the trigger rate was not matching
the trigger source.
Find out that Link 3 was not collecting and |
|
10
|
06 Feb 2019 14:12 |
Pierre | Configuration | Hardware | Extended Trigger Time Tag (ETTT) | Confirmed this ETTT configuration is working.
ETTT Enabled [22..21] |
|
14
|
19 Feb 2019 16:26 |
Pierre-A. | Configuration | General | Overall HW configuration | For Reference,
I put a simple schematic for the
Trigger/Run control. |
|
22
|
27 Mar 2019 08:57 |
Pierre | Configuration | Hardware | CERN setup | Found that the Trigger out from the CB is
on output1
Trigger / Not used
|
|
23
|
27 Mar 2019 14:04 |
Pierre | Configuration | Trigger | Time stamp sync | The ChronoBox latest FW is loaded. Let tme
know if this is what the chronobox should
look like in term of registers.
|
|
24
|
28 Mar 2019 02:18 |
Pierre | Configuration | Trigger | Test | |
|
32
|
12 Jul 2019 01:27 |
Marco Rescigno | Configuration | | MB1 test in proto-0 setup/1 | Changed custom size to 500 (20 us), tested
ok run 681 |
|
47
|
27 Oct 2019 06:08 |
Simone Stracka | Configuration | Hardware | Converters installed in VME crate | People: Edgar, Simone
Installed differential to single-ended
converters in VME crate, and turned crate |
|
52
|
28 Oct 2019 15:10 |
Simone Stracka | Configuration | Hardware | LV for steering module and current status of CAEN mainframe | The HV board from Naples did not turn on:
Yury gave it to the CAEN guys to check and/or
bring back for replacement. |
|
55
|
01 Nov 2019 06:50 |
Alex Kish | Configuration | Hardware | Turn off the fields | Ramp down the fields, in 200V steps.
Nominal values:
1st ring: 4180 V
|
|
58
|
01 Nov 2019 09:49 |
Alex Kish | Configuration | Hardware | Turn off the fields | Ramp UP the fields, in 100V steps.
Nominal values:
|
|
65
|
04 Nov 2019 06:02 |
Alex Kish | Configuration | Hardware | Increased fields! | Increased drift and extraction fields.
Current settings: wire gate 5100 V, 1st ring: |
|
66
|
04 Nov 2019 07:57 |
Ben Smith | Configuration | Software | Package installations | This morning I installed the xrootd-client
and python3-devel packages from yum. For
the latter to work, I needed to do a yum |
|
67
|
04 Nov 2019 08:00 |
Ben Smith | Configuration | MIDAS | Change of location of history files | /home was getting very full on ds-proto-daq,
so I moved the history files onto the /data
disk. There is now a symlink from ~/online/history |
|
70
|
05 Nov 2019 01:22 |
Ben Smith | Configuration | Digitizer | V1725 board config | The V1725s have been changed to have board
config 0x50 rather than 0x10. This means
that they will now trigger on the leading |
|