HW-Time-Slice Firmware-Filter Software vx2740 Steering Module Reports MVM MVM Vexos MVM-Bug listing MVM TRIUMF Local DS Prototype DS Cryogenic For Shifters BCIT-31 ChronoBox Run Operation DS-DAQ
  CERN DS-Proto0 read-only backup  Not logged in ELOG logo
Message ID: 10     Entry time: 06 Feb 2019 14:12
Author: Pierre 
Type: Configuration 
Category: Hardware 
Subject: Extended Trigger Time Tag (ETTT) 
Confirmed this ETTT configuration is working.

                     ETTT Enabled [22..21] = b10
                      |
                      v
Data [0x811C] = 0x 00 4 D 013C

                ETTT Time [47..32]
                 |  Ch Mask[16..0]  Time[31..0]
                 |    TTTTTT           |
Header 1         v    v    v           v
0xa0001914 0x00 0025 ff 0xff1ca598 0xe9c6e8a1   < event 1
0xa0001914 0x00 0025 ff 0xff1ca599 0xe9c82e25   < event 2

dTime :  0x25e9c82e25 &#8722; 0x25e9c6e8a1 = 0x14584 => 83332
Time interval: 8ns  => 666.7e-6s => Freq: 1500Hz corresponding the current trigger rate

PAA