51
|
28 Oct 2019 14:56 |
Simone Stracka | Problem | Software | Issues with network configuration for steering module |
We have tried to connect the steering module to the DAQ pc via Luigi's USB-Ethernet adapter.
That does not appear to work. We'll receive another USB-Ethernet adapter tomorrow, and it should be configured to have a static IP address : 192.168.121.1 and NetMask 255.255.255.0
For the time being, we are using Tom's laptop.
|
52
|
28 Oct 2019 15:10 |
Simone Stracka | Configuration | Hardware | LV for steering module and current status of CAEN mainframe |
The HV board from Naples did not turn on: Yury gave it to the CAEN guys to check and/or bring back for replacement.
Yi and Luigi rented a new HV module (A1519). The HV module (A1520P) we used for tests of the I-V script is also present in the Mainframe (see below).
The first 24 channels (slots 0 and 1) are therefore HV. If the new HV does not show up on time we'll try and adapt the cables to work with A1519.
The A2517A module is LV. This is currently operated from the DAQ pc using CAEN_HVPSS_ChannelsController.jnlp (located in the Desktop/SteeringModule folder).
The three low voltage channels (0,1,2) should be turned on at the same time by setting Pw = ON.
Settings:
Channel 0 and 1: I0Set = 2.0 A , V0Set = 2.5 V , UNVThr = 0 V, OVVThr = 3.0 V, Intck = Disabled
IMon = 1.44 (this depends on the illumination) , VMon = 2.48 V , VCon = 2.79 V
Channel 2: I0Set = 1.0 A , V0Set = 5.0 V , UNVThr = 0 V, OVVThr = 5.5 V, Intck = Disabled
IMon = 0.07 , VMon = 4.998 V , VCon = 5.02 V
In case the channels trip they cannot be ramped back up unless the alarms are cleared.
|
Attachment 1: CAENmainframe.png
|
|
124
|
12 Nov 2019 10:08 |
Edgar Sanchez | Routine | General | Laser run |
Run 1096 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1097 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Problem: The access to the Chronobox page is not possible. Rate 400 Hz |
134
|
15 Nov 2019 00:19 |
Yi Wang | Routine | General | Laser run |
Run 1166 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Trigger rate went to 0, run was stopped manually.
Run 1169 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1170 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
|
140
|
18 Nov 2019 02:40 |
Pascal/Julie | Routine | General | Laser run |
Run 1232 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1233 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k |
144
|
19 Nov 2019 00:14 |
Pascal/Julie | Routine | General | Laser run |
Run 1252 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Low trigger rate - must restart manually
Run 1253 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1254 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k |
151
|
14 Oct 2020 08:39 |
Tom Thorpe | Routine | General | Log regarding MB2 removal from Proto-0 |
Drive folder with data files and some pictures: https://drive.google.com/drive/folders/1Sr27kQTOO2kciBY7v-wYBP06tWLVEjiP
Lab notes @ CERN
11/10/2020
11:00 – First look at MB2. Noticed on some of the PDMs that some of the tile PCB are more delaminated than others. PDM 25 also had the FEB connectors more inserted than any other PDMs. No other differences noticed.
11:30 – Connected SMU and switching matrix to MB2. Power w/o steering module. Checked that all channels match by running photo current tests with and w/o a flashlight on one corner. Using the computer form Pisa with Labview scripts from Matteo. Everything is ok.
12:00 – Replace plastic shields around Proto-0 for IV curve measurements in dark conditions. PDM 1 breakdown is around 63V. Looks ok. PDM 2 is over 65V…. Seems high.
12:30 - PDM 6 People walking, disrupted the light conditions. IV curve will be jagged. Will go for lunch. Closing door and turning off exterior lights. Current will decrease
15:30 – IV curves look fine. …/MB2_retriveval_CERN/data_files/iv_warm_1
16:00 – Took forward IV curves. Saved as …/MB2_retriveval_CERN/data_files/forward_warm_1. They look fine. Checked old pictures of the lamination of the tile PCB and they have always varied in thickness, from before the LNGS LN test.
16:45 – Power FEBs on to look at noise spectra with 20V bias on the tiles.
18:00 – PDM 2, 4, 6, and 25 have very odd power spectra. George said they look like they’re not connected.
18:30 – Saving FFT to disk for analysis. Saved as …/MB2_retriveval_CERN/data_files/noise_warm_1
19:00 – PDM 2,4,6 and 25 have one of the differential lines disconnected. Using a light and looking at the differential outputs on the scope, this is obvious. 20V bias and FEBs powered at 5V. Saved as …/MB2_retriveval_CERN/data_files/waveform_light_1
19:30 – Started another round of reverse IV curves in dark conditions. Saved as …/MB2_retriveval_CERN/data_files/iv_warm_2
21:15 – IV still going. EP will come back to shut off the LV supplies
12/12/2020
10:00 – Checking connectivity of all the differential lines. One side of PDM 6 is not connected. 2, 4, and 25 are connected. Doesn’t explain what we saw yesterday. Channel 2 and 4 now are responding to light on both differential lines. The only thing that has changed is unplugging the external signal cable and plugging it back in. PDM 6 is the only one now not responding on both lines.
10:30 – Checking the power spectra. PDM 4 and 6 are bad on both differential lines. PDM 2 looks fine. Seems to be non-repeatable.
11:00 – Disconnected the signal patch board from the finger strip. Holding the patch and reconnecting the entire signal path and pinging now shows PDM 6 and 25 have one differential line disconnected.
11:30 – Checking the power spectra on every PDM channel directly from the fingerstrip with a flat cable.
12:30 – PDMs 2, 4, 6 power spectra from the fingerstrip look normal. Both connections are there. Light response is normal. Channel 13, however, now has one differential line disconnected and the other looks abnormal. Saving waveform data with 60Hz light on as waveform_light_2.
All of this is pointing to multiple connection issues…
15:00 – It looks like PDM 13 is disconnected from the fingerstrip, or at least it is pulled further away than the others. This is likely also due to the removal of the patch. PDM 2 also looks pulled a bit further.
15:30 – Unmounted MB2 and put on the stand to do an optical inspection. The microscope didn’t allow a close enough look at the wirebonds to search for any broken ones. A close look by eye didn’t reveal any obvious damage.
16:00 – Packed everything up. MB2 needs a closer look, likely needs to be disassembled. |
Attachment 1: forward_warm_1.png
|
|
Attachment 2: iv_warm_1.png
|
|
Attachment 3: iv_warm_2.png
|
|
37
|
13 Jul 2019 14:50 |
Marco Rescigno | Routine | General | Long Laser run on disk |
Run 747 , 250 k events
as requested by alessandro 40% post trigger, acquisition window 16 us.
|
32
|
12 Jul 2019 01:27 |
Marco Rescigno | Configuration | | MB1 test in proto-0 setup/1 |
Changed custom size to 500 (20 us), tested ok run 681 |
33
|
12 Jul 2019 02:05 |
Marco Rescigno | Problem | Trigger | MB1 test in proto-0 setup/1 |
Tried to get laser sync signal into chronobox (clkin1 input), apparently all triggers dropped by chronobox
Problem is that the same now happen also with the regular setup where the clkin1 signal is taken by the dual timer.
|
36
|
12 Jul 2019 11:34 |
Marco Rescigno | Routine | General | MB1 test in proto-0 setup/2 Run 718 |
Run 718 has 25 channel readout Board 0 ch 0-15 and Board 0 ch 0-7.
Laser triggered, some noise, maybe not so bad as is look here.
RMS of baseline is 200 ADC count. Single PE peak is about 100-200 ADC count from baseline.
Looking carefully, some very nice scintillation event are found.
|
Attachment 1: RMS_run718.gif
|
|
Attachment 2: SPE.gif
|
|
Attachment 3: Scintillation.gif
|
|
35
|
12 Jul 2019 07:58 |
Marco Rescigno | Routine | General | MB1 test in proto-0 setup/day 2 |
Implemented a simple charge integration on the midas display, most of the channels look almost as good as this one.
Nothing on the daq side changed, a part from the DAC value moved to 5000, allowing a slightly greater dynamic range. |
Attachment 1: Screenshot_from_2019-07-13_17-45-14.png
|
|
45
|
25 Oct 2019 07:44 |
Simone Stracka | Routine | Hardware | MB2 activities on October 25 |
Crew: Edgar Sanchez Garcia, Sam Hill, Simone Stracka, Tom Thorpe, Yi Wang
Morning: Set up equipment for the IV curve measurements. We enable the interlock system in the SMU in order to go higher than 42 V (handle with care).
Afternoon: Measuring IV curves. Note there is a voltage drop in the LV cable that needs to be compensated for.
Chan 1 is set at +3.3V drawing 1.46A; sense line reads +2.39V
Chan 2 is set at -3.3V drawing 1.46A; sense line reads -2.39V
IV curves are scanning from 0-80V in 200mV steps w/ 100uA limit: elog:45/2 , elog:45/3
The two runs are visible ( elog:45/1 for color code and mapping )
Compare to elog:45/4 , obtained when probing individual tiles under more controlled light conditions |
Attachment 1: TileFEBMapping_MB2.png
|
|
Attachment 2: iv_warm_1.png
|
|
Attachment 3: iv_warm_1_zoom.png
|
|
Attachment 4: IndividualTileAtLNGS.png
|
|
30
|
02 Jul 2019 18:01 |
Thomas | Routine | Software | MIDAS running again |
Darkside people seem to be doing some tests at CERN next week. It looks like they aren't going to use our DAQ (they will I think just use CAEN tools). But we took
opportunity to make TRIUMF DAQ work again. There was a couple issues
1) I needed to inver the busy signal for input 0 to the chronobox in order to get any triggers out of the system. Not sure why; I don't think I had to do that before. But it is
working now.
2) The CERN web proxy was not initially working. Somehow I 'fixed' it by changing the proxy configuration variable of SERVICE_NAME from 'midas' to 'dummy'. I don't
understand why that fixed it. But now you can see the DAQ here:
https://m-darkside.web.cern.ch/chronobox/ |
29
|
11 Apr 2019 08:20 |
Thomas | Routine | Trigger | Missing ZMQ banks |
I have done a couple longer tests of the DAQ setup. The runs were done with a high trigger rate of ~60Hz, with
the V1725 asserting their busy to throttle the trigger.
I found that after a couple hours (~5hours) that we would no longer be getting ZMQ packets from the chronobox.
You can see this with error message like
15:29:04.092 2019/04/11 [feov1725MTI00,ERROR] [feoV1725.cxx:1033:read_trigger_event,ERROR] Error: did not
receive a ZMQ bank. Stopping run.
10:26:37.158 2019/04/11 [mhttpd,INFO] Run #657 started
If I start a new run I am still missing ZMQ packets. However, if I restart the frontend program (and hence
re-initialize the ZMQ link), then the chronobox does start sending triggers again. So it may be more of a
problem with the ZMQ setup in the MIDAS frontend. Needs more investigation. |
101
|
08 Nov 2019 10:07 |
Edgar Sanchez | Routine | General | Monitoring Data Taking |
Run 995
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 230 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz
Comment: TPC pressure increased up to 290 mbarg
Run 996
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 225 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz
Comment: TPC pressure increased up to 240 mbarg
Run 997
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz
Run 998
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz |
104
|
09 Nov 2019 00:13 |
Xiang Xiao | Routine | General | Morning data taking |
"ERROR INFO: 10:38:10.924 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."
Emailed to Ben.
Run 999
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 233 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1000
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 235 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1001
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 238 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1002
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 238 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1003 Laser run (manually stopped)
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Comment: Triger rate too low (~8 Hz), manually stopped
Run 1004 Laser run (manually stopped)
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Comment: Triger rate too low (~8 Hz), manually stopped
|
115
|
11 Nov 2019 01:12 |
Yi / Edgar | Routine | General | Morning data taking |
Run 1065
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 100us total, 80 us post-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1066
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Trace length: 100us total, 80 us post-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1067
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Trace length: 100us total, 80 us post-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 20k |
121
|
12 Nov 2019 00:24 |
Yi Wang | Routine | General | Morning data taking |
Run 1084
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1085
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1086
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 6k
Run crashed: 09:31:55.239 2019/11/12 [feov1725MTI00,ERROR] [feoV1725.cxx:663:link_thread,ERROR] Exiting thread 0 with error |
125
|
13 Nov 2019 00:08 |
Yi Wang | Routine | General | Morning data taking |
Run 1099
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 250 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1100
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 250 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1101
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 250 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1102
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 250 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1103
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADCc below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 250 mbarg
SiPM HV: 65 V
Number of events: 10k |