Message ID: 36     Entry time: 12 Jul 2019 11:34
Author: Marco Rescigno 
Type: Routine 
Category: General 
Subject: MB1 test in proto-0 setup/2 Run 718 

Run 718 has 25 channel readout Board 0 ch 0-15 and Board 0 ch 0-7. 

Laser triggered, some noise, maybe not so bad as is look here.

RMS of baseline is 200 ADC count. Single PE peak is about 100-200 ADC count from baseline.

Looking carefully, some very nice scintillation event are found.

 

 

 

 