151
|
14 Oct 2020 08:39 |
Tom Thorpe | Routine | General | Log regarding MB2 removal from Proto-0 | Drive folder with data files and some pictures: https://drive.google.com/drive/folders/1Sr27kQTOO2kciBY7v-wYBP06tWLVEjiP
Lab notes @ CERN
11/10/2020
11:00 – First look at MB2. Noticed on some of the PDMs that some of the tile PCB are more delaminated than others. PDM 25 also had the FEB connectors more inserted than any other PDMs. No other differences noticed.
11:30 – Connected SMU and switching matrix to MB2. Power w/o steering module. Checked that all channels match by running photo current tests with and w/o a flashlight on one corner. Using the computer form Pisa with Labview scripts from Matteo. Everything is ok.
12:00 – Replace plastic shields around Proto-0 for IV curve measurements in dark conditions. PDM 1 breakdown is around 63V. Looks ok. PDM 2 is over 65V…. Seems high.
12:30 - PDM 6 People walking, disrupted the light conditions. IV curve will be jagged. Will go for lunch. Closing door and turning off exterior lights. Current will decrease
15:30 – IV curves look fine. …/MB2_retriveval_CERN/data_files/iv_warm_1
16:00 – Took forward IV curves. Saved as …/MB2_retriveval_CERN/data_files/forward_warm_1. They look fine. Checked old pictures of the lamination of the tile PCB and they have always varied in thickness, from before the LNGS LN test.
16:45 – Power FEBs on to look at noise spectra with 20V bias on the tiles.
18:00 – PDM 2, 4, 6, and 25 have very odd power spectra. George said they look like they’re not connected.
18:30 – Saving FFT to disk for analysis. Saved as …/MB2_retriveval_CERN/data_files/noise_warm_1
19:00 – PDM 2,4,6 and 25 have one of the differential lines disconnected. Using a light and looking at the differential outputs on the scope, this is obvious. 20V bias and FEBs powered at 5V. Saved as …/MB2_retriveval_CERN/data_files/waveform_light_1
19:30 – Started another round of reverse IV curves in dark conditions. Saved as …/MB2_retriveval_CERN/data_files/iv_warm_2
21:15 – IV still going. EP will come back to shut off the LV supplies
12/12/2020
10:00 – Checking connectivity of all the differential lines. One side of PDM 6 is not connected. 2, 4, and 25 are connected. Doesn’t explain what we saw yesterday. Channel 2 and 4 now are responding to light on both differential lines. The only thing that has changed is unplugging the external signal cable and plugging it back in. PDM 6 is the only one now not responding on both lines.
10:30 – Checking the power spectra. PDM 4 and 6 are bad on both differential lines. PDM 2 looks fine. Seems to be non-repeatable.
11:00 – Disconnected the signal patch board from the finger strip. Holding the patch and reconnecting the entire signal path and pinging now shows PDM 6 and 25 have one differential line disconnected.
11:30 – Checking the power spectra on every PDM channel directly from the fingerstrip with a flat cable.
12:30 – PDMs 2, 4, 6 power spectra from the fingerstrip look normal. Both connections are there. Light response is normal. Channel 13, however, now has one differential line disconnected and the other looks abnormal. Saving waveform data with 60Hz light on as waveform_light_2.
All of this is pointing to multiple connection issues…
15:00 – It looks like PDM 13 is disconnected from the fingerstrip, or at least it is pulled further away than the others. This is likely also due to the removal of the patch. PDM 2 also looks pulled a bit further.
15:30 – Unmounted MB2 and put on the stand to do an optical inspection. The microscope didn’t allow a close enough look at the wirebonds to search for any broken ones. A close look by eye didn’t reveal any obvious damage.
16:00 – Packed everything up. MB2 needs a closer look, likely needs to be disassembled. |
Attachment 1: forward_warm_1.png
|
|
Attachment 2: iv_warm_1.png
|
|
Attachment 3: iv_warm_2.png
|
|
150
|
08 Sep 2020 12:46 |
Ben Smith | Configuration | Other | New backup ELOG location | I have added a backup of the CERN proto-0 ELOG at TRIUMF, in case ds-proto-daq is unavailable in future.
The backup ELOG is at https://ladd00.triumf.ca/elog-ds/. There are other logbooks hosted in the same place (requiring a username/password), but the CERN backup logbooks (at the bottom of the list) are available to all. I have configured things so that the TRIUMF replicas are read-only. To add new entries one must still use the CERN logbooks.
The logbooks are rsynced every hour via cron. |
149
|
16 Mar 2020 14:21 |
Ben Smith | Configuration | Software | New CAEN HV frontend | A few months ago I created a new midas frontend for CAEN HV crates. It behaves more like what users expect - changes made through other interfaces (e.g. ssh or the Java-based GUI) will be reflected in the ODB. The ODB structure is created dynamically based on which modules are present in the crate, and the parameters that each module supports (e.g. whether you specify a ramp-up rate in V/s or a ramp-up time in s).
When I first wrote the frontend, ds-proto-daq was offline. Pierre reminded me today that I should actually deploy the new code.
The new program code is at ~/online/dsproto_sy4527/caen_hv*. The executable is at ~/online/bin/caenhv_fe. The frontend is available on the "Programs" page in mhttpd. If we want to run the old HV frontend, it is still available at ~/online/bin/sy4527.
Connection parameters for the frontend are found in the ODB at /Equipment/CAEN_HV/Settings/Global. Currently the crate appears to be off, so the frontend will refuse to start properly as it cannot connect to 192.168.1.9. If the frontend is started with the crate on, it will create more directories in /Equipment/CAEN_HV/Settings (one for each module). The "Voltage" page on the experiment now points to a generic webpage that will work with arbitrary combinations of modules. |
148
|
20 Nov 2019 04:47 |
Sam Hill | Routine | General | IV Curve - warming up | Temp at start:
A: 242.7K B: 278.16K C: 86.4K D: 240.89K
Temp at End:
A: 248.87K B: 278.24K C: 86.37K D: 247.75K
Took iv curves, data is on DAQ PC here: /home/dsproto/online/dsproto_sy4527/ivdata_191120 |
147
|
19 Nov 2019 04:07 |
Julie / Pascal | Routine | Other | IV curve at the end of data taking | IV of 25 SiPM put in ivdata_191119 directory |
146
|
19 Nov 2019 01:39 |
Julie / Pascal | Routine | General | Random Trigger runs suggested by Simone, S1 data / 3ms | Run 1258 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 3000us, 1500us pre-trigger
No field
SIPM HV: 65V
Number of events: 10k
--> stopped after 4.7k (end of the data taking at 11:00) |
145
|
19 Nov 2019 00:57 |
Julie/ Pascal | Routine | General | Random Trigger runs suggested by Simone, S1 data / 1ms | Run 1255 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 1000us, 500us pre-trigger
No field
SIPM HV: 65V
Number of events: 10k
Run 1256 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 1000us, 500us pre-trigger
No field
SIPM HV: 65V
Number of events: 10k
--> Stop after 1.7k evts (10:16)
Run 1257 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 1000us, 500us pre-trigger
No field
SIPM HV: 65V
Number of events: 10k |
144
|
19 Nov 2019 00:14 |
Pascal/Julie | Routine | General | Laser run | Run 1252 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Low trigger rate - must restart manually
Run 1253 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1254 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k |
143
|
18 Nov 2019 07:20 |
Ben Smith | Routine | General | Disk cleanup | Lazy logger has caught up with the backlog, and continues to transfer new files to darkside01 as they are written.
EOS currently contains data for runs 822-1201.
I have deleted the following runs from ds-proto-daq:
* runs < 700 that were taken with gzip compression
* runs 800-1100 that are available on EOS and darkside01
The following files remain:
* runs 700-800 that haven't been copied to EOS (but are on darkside01). If they're of interest, they should be copied to EOS.
* runs 1200+ that haven't been copied to EOS yet (but will be soon).
Disk usage has improved from 85% to 33%. |
142
|
18 Nov 2019 07:01 |
Julie / Edgar | Routine | General | Random Trigger runs suggested by Simone, S1 data | PDM 15 and PDM 19 strange behaviour, lot of noise
Run 1239 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1240 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1241 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1242 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1243 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1244 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1245 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1246 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1247 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1248 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1249 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1250 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1251 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
Run 1252 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
No field
SIPM HV: 65V
Number of events: 40k
|
141
|
18 Nov 2019 03:03 |
Julie / Pascal / Edgar | Routine | General | Random Trigger runs suggested by Simone (4 runs ~ 200 k evts) | Run 1234 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
Drift field 200 V/cm Extraction 3.78 kV/cm
SIPM HV: 65V
Number of events: 200k
Problem with DAQ at 40k, force to restart manually
Run 1235 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
Drift field 200 V/cm Extraction 3.78 kV/cm
SIPM HV: 65V
Number of events: 200k
Problem with DAQ at 70k, force to restart manually
Run 1237 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
Drift field 200 V/cm Extraction 3.78 kV/cm
SIPM HV: 65V
Number of events: 40k
Run 1238 Physics run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Threshold: external clock, set to 50 Hz
Trace length: 200us, 100us pre-trigger
Drift field 200 V/cm Extraction 3.78 kV/cm
SIPM HV: 65V
Number of events: 40k
|
140
|
18 Nov 2019 02:40 |
Pascal/Julie | Routine | General | Laser run | Run 1232 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1233 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k |
139
|
18 Nov 2019 00:18 |
Yi / Edgar | Routine | General | dual phase data taking | We still have the noise issue.
Run 1219
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1220
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1221
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1222
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1223
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1224
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1226
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 40k
Run 1228
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 15k
Run 1229
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 40k |
138
|
17 Nov 2019 02:23 |
Yi Wang | Routine | General | dual phase data taking | Noise problem seems like fixed after remaking the Al foil cover.
Run 1208
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1209
Gas pocket: ON (thinkness 9mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
At the end of run 1209, a lot of noise show up again. |
137
|
17 Nov 2019 00:48 |
Yi Wang | Routine | General | dual phase data taking | Run 1197
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1198
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1200
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1201
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 3
Trace length: 100us total, 50 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 30 mbarg
SiPM HV: 65 V
Number of events: 20k
Data taking has to be stopped because of the noises.
|
Attachment 1: Screenshot_from_2019-11-17_11-02-26.png
|
|
136
|
16 Nov 2019 01:57 |
Yi Wang | Routine | General | Dual phase data taking proposed by Marco | Run 1186
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 100 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 10k
V1725 offline, run stopped manually
Run 1188
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 100 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 70k
Trigger rate became very low, run stopped manually
Run 1189
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1190
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 10
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1191
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 200 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 15
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1192
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1193
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 10
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1194
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 2
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 15
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1195
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 500 ADCc below baseline
Coincidence: 2
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5
TPC Pressure: 26 mbarg
SiPM HV: 65 V
Number of events: 10k |
135
|
15 Nov 2019 10:48 |
E. Sanchez | Routine | General | S1 data taking proposed by Marco | Run 1171
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1172 Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1173
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 4 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1174
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns TPC
Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1175
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 6 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1176
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 500 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1177
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 500 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1178
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 500 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 4 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1179 Gas pocket: OFF
Fields: 0 V/cm
Threshold: 500 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1180
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 500 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 6 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1181
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 200us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 100k
Problem with DAQ, forced to stop
Run 1182
Gas pocket: OFF
Fields: 0 V/cm
Threshold: 200 ADCc below baseline
Trace length: 200us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 80 mbarg
SiPM HV: 65 V
Number of events: 100k |
134
|
15 Nov 2019 00:19 |
Yi Wang | Routine | General | Laser run | Run 1166 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Trigger rate went to 0, run was stopped manually.
Run 1169 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1170 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock, laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
|
133
|
14 Nov 2019 08:14 |
Ben Smith | Routine | General | Automatic copying to darkside01 | I have set up automatic copying of data from ds-proto-daq to darkside01, using the lazy logger.
There are 2 lazy logger channels set up - one for lz4 files and one for gz files. It will take some time for all the files to copied over. Note that lazy logger writes a message into the midas log for every file that it copies. The lazy logger channels can be started from the Program page like all the other midas-related programs.
Once files have been copied to darkside01 (by lazy logger) and to EOS (by Pablo/Edgar) we can delete files from ds-proto-daq. We will probably not be able to do this until tomorrow (copying is quite slow). |
132
|
14 Nov 2019 07:42 |
E. Sanchez | Routine | General | S1 data taking with field | Run 1148
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
problem DAQ, forced to stop manually
Run 1149
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 380 Hz
Run 1150
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 310 Hz
Run 1151
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 400 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 250 Hz
Run 1152
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 600 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 150 Hz
Run 1153
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 1000 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 5 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 90 Hz
Run 1154
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 390 Hz
Run 1155
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
Run 1156
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 400 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
Run 1157
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 600 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
Run 1158
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 1000 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 2 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
Run 1159
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Run 1160
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Run 1161
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 400 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Run 1163
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 600 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Run 1164
Gas pocket: OFF
Fields: 500 V/cm
Threshold: 1000 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Coincidence: 3 PDMs
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k |
|