5
11 Jan 2019 11:07
Thomas
Routine
Digitizer
Data corruption for ADC channels
I modified the analyzer so that it shows data
from all four modules.
7
21 Jan 2019 13:05
Thomas, Pierre
Routine
Digitizer
Tests of V1725 baseline and triggering
A little progress on a couple fronts:
1) We got the baseline DAC setting working
15
05 Mar 2019 10:36
Thomas
Routine
Digitizer
Switched to standard V1725 firmware
It turns out that the ZLE V1725 firmware we
are using only supports reading out up to
4000 samples per channel.
70
05 Nov 2019 01:22
Ben Smith
Configuration
Digitizer
V1725 board config
The V1725s have been changed to have board
config 0x50 rather than 0x10. This means
that they will now trigger on the leading
114
11 Nov 2019 00:47
Yi / Edgar
Problem Fixed
Digitizer
Problem fixed by restarting the VME crate
119
12 Nov 2019 00:16
Yi Wang
Problem
Digitizer
V1725 error, runs keep crashing
09:15:33.955 2019/11/12 [feov1725MTI00,ERROR]
[feoV1725.cxx:663:link_thread,ERROR] Exiting
thread 3 with error
120
12 Nov 2019 00:23
Ben Smith
Problem
Digitizer
V1725 error, runs keep crashing
It looks like there's a memory leak in the
high voltage driver, and we were running
out of memory (using a lot of swap). That
32
12 Jul 2019 01:27
Marco Rescigno
Configuration
MB1 test in proto-0 setup/1
Changed custom size to 500 (20 us), tested
ok run 681