ID Date Author Type Categoryup Subject
  83   05 Nov 2019 08:14 Alex / SamRoutineGeneralRun 898, Edrift 200V/cm

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar) [Edgar\Yi: correction: OFF (during recirculation tpc pressure about 400 mbar)]

Fields: drift 200 V/cm, extraction 2.8 kV/cm (wire grid 3780 V, 1st ring 4180 V, cathode 6180 V) 

Threshold: 400 ADC below baseline (baselines equalized to 15500)

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment:

  84   05 Nov 2019 08:23 Alex / SamRoutineGeneralRun 899, Edrift 100V/cm

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar) [Edgar\Yi: correction: OFF (during recirculation tpc pressure about 400 mbar)]

Fields: drift 100 V/cm, extraction 2.8 kV/cm (wire grid 3780 V, 1st ring 3580 V, cathode 3780 V) [Yi: 1st ring setting is incorrect, inverted fied between wire grid and field cage ring]

Threshold: 400 ADC below baseline (baselines equalized to 15500)

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment:

  85   05 Nov 2019 08:32 Alex / SamRoutineGeneralRun 900, Edrift 500V/cm

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar) [Edgar\Yi: correction: OFF (during recirculation tpc pressure about 400 mbar)]

Fields: drift 500 V/cm, extraction 2.8 kV/cm (wire grid 3780 V, 1st ring 2780 V, cathode 7780 V) [Yi: 1st ring setting is incorrect, inverted fied between wire grid and field cage ring]

Threshold: 400 ADC below baseline (baselines equalized to 15500)

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment:

  86   05 Nov 2019 23:58 Yi WangRoutineGeneralMorning data taking, purity check

Run 917

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 918

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 919

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 920

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

  88   06 Nov 2019 11:22 Yi WangRoutineGeneralEvening data taking, purity check

Not sure if the trigger threshold is correctly set, there is an error:

Traceback (most recent call last):
  File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 71, in <module>
    main()
  File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 47, in main
    (mean, rms) = fit_res
ValueError: too many values to unpack (expected 2)

the trigger threshold for all the runs in this Elog is 2699 ADCc below the baseline (see Elog 89)
 

Run 943

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 945

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 946

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 948

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 949

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 951

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 952

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 954

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

  89   06 Nov 2019 12:28 Ben SmithProblem FixedGeneralEvening data taking, purity check
> Not sure if the trigger threshold is correctly set, there is an error:

Sorry, I made a change to the baseline calculation and forgot to update the threshold tool. It's updated now.

The thresholds currently seem to be set to 12801 - this is 2699 below the nominal baseline of 15500.
  90   07 Nov 2019 02:07 SamRoutineGeneralRun 963 Laser Run with SiPMs at 65V, 12us trace

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

  91   07 Nov 2019 02:07 Yi/Marco/AndreaRoutineGeneralmorning data taking, purity check

Run 956

Gas pocket: ON (thickness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 957

Gas pocket: ON (thickness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 958

Gas pocket: ON (thickness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 960

Gas pocket: ON (thickness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 961

Gas pocket: ON (thickness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

  92   07 Nov 2019 02:16 SamRoutineGeneralRun 964 Laser Run with SiPMs at 65V, 30us trace

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 30us, 5us pre-trigger

SIPM HV: 65V

  93   07 Nov 2019 02:34 SamRoutineGeneralRun 966 Laser Run with SiPMs at 65V, 30us trace

Laser Intensity: MIN 

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 30us, 5us pre-trigger

SIPM HV: 65V

 

MISSING FILE (E.Sanchez)

  94   07 Nov 2019 23:48 Yi WangRoutineGeneralmorning data taking

Run 973

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 974

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 975

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 976

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 978 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 979 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

After laser run:

Switch Laser OFF

In Chronobox:

DIsable "External Triger Enable [ext_trig_ena]0\[0]

Input "0x3F3F3F7F" in "Enable Channel [ch_enable] 

 

  101   08 Nov 2019 10:07 Edgar SanchezRoutineGeneralMonitoring Data Taking

Run 995

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  230 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

Comment: TPC pressure increased up to 290 mbarg

Run 996

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  225 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

Comment: TPC pressure increased up to 240 mbarg

Run 997

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

Run 998

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 10k

Rate  ~9 Hz

  104   09 Nov 2019 00:13 Xiang XiaoRoutineGeneralMorning data taking

"ERROR INFO: 10:38:10.924 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."

Emailed to Ben.

 

Run 999

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  233 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1000

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  235 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1001

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  238 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1002

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  238 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1003 Laser run (manually stopped)

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Comment: Triger rate too low (~8 Hz), manually stopped

 

Run 1004 Laser run (manually stopped)

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Comment: Triger rate too low (~8 Hz), manually stopped

 

  106   09 Nov 2019 09:39 Sam HillRoutineGeneralEvening data taking

Run 1016

Gas pocket: ON (thinkness 7mm)

Fields: Off

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: 201 mbarg

SiPM HV: 65 V

Number of Events: 10K

 

Run 1017

Gas pocket: ON (thinkness 7mm)

Fields:  drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: 201 mbarg

SiPM HV: 65 V

Number of Events: 10K

 

Run 1018

Gas pocket: ON (thinkness 7mm)

Fields:  drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: 217 mbarg

SiPM HV: 65 V

Number of Events: 10K

 

Run 1019

Gas pocket: ON (thinkness 7mm)

Fields:  drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: 240

SiPM HV: 65 V

Number of Events: 10K

 

Run 1020 Laser Run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Comment: Recieved this error message but run seemed to continue ok and data appeared to be written: 19:56:04.839 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run.   â•³

Run 1021 Laser Run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

 

 

 

  107   10 Nov 2019 00:06 Yi WangRoutineGeneralmorning data taking

Run 1022

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  159 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1024

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  186 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1025

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  207 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1026

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  225 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1028 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 1029 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

  108   10 Nov 2019 08:04 Edgar SanchezRoutineGeneralS1 data taking with field

Run 1032

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 800 Hz

Run 1033

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 350 Hz

Run 1034

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 250 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 260 Hz

Run 1035

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 200 Hz

Run 1036

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 130 Hz

 

"ERROR INFO: 17:19:23.176 2019/11/10 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."

Run 1037

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 800 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  110 Hz

Run 1038

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  80 Hz

Run 1039

Gas pocket: OFF

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 150 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

 

  109   10 Nov 2019 08:40 Edgar SanchezRoutineGeneralS1 data taking without field

Run 1041

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 360 Hz

Run 1042

Gas pocket: OFF

Fields: OFF

Threshold: 200 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 340 Hz

Run 1043

Gas pocket: OFF

Fields: OFF

Threshold: 250 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 290 Hz

Run 1044

Gas pocket: OFF

Fields: OFF

Threshold: 400 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 240 Hz

Run 1045

Gas pocket: OFF

Fields: OFF

Threshold: 600 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 170 Hz

Run 1046

Gas pocket: OFF

Fields: OFF

Threshold: 800 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  140 Hz

Run 1047

Gas pocket: OFF

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  110 Hz

Run 1048

Gas pocket: OFF

Fields: OFF

Threshold: 150 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

Run 1049

Gas pocket: OFF

Fields: OFF

Threshold: 100 ADCc below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  340 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate:  400 Hz

  110   10 Nov 2019 11:18 Edgar SanchezRoutineGeneralAfternoon system monitoring

Run 1050

Gas pocket: ON (unknow)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  245 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1051

Gas pocket: ON (unknow)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 10k

Run 1052

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  245 mbarg

SiPM HV: 65 V

Number of events: 10k

Stopped. Problem with DAQ

  111   10 Nov 2019 11:19 Edgar SanchezRoutineGeneralS1 + S2 short time window

Run 1054

Gas pocket: ON (unknow)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 100 ADC below baseline

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 50k

Rate: 400 Hz 

 

 

  115   11 Nov 2019 01:12 Yi / EdgarRoutineGeneralMorning data taking

Run 1065

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1066

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1067

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k