ID |
Date |
Author |
Type |
Category |
Subject |
83
|
05 Nov 2019 08:14 |
Alex / Sam | Routine | General | Run 898, Edrift 200V/cm | Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar) [Edgar\Yi: correction: OFF (during recirculation tpc pressure about 400 mbar)]
Fields: drift 200 V/cm, extraction 2.8 kV/cm (wire grid 3780 V, 1st ring 4180 V, cathode 6180 V)
Threshold: 400 ADC below baseline (baselines equalized to 15500)
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Comment: |
84
|
05 Nov 2019 08:23 |
Alex / Sam | Routine | General | Run 899, Edrift 100V/cm | Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar) [Edgar\Yi: correction: OFF (during recirculation tpc pressure about 400 mbar)]
Fields: drift 100 V/cm, extraction 2.8 kV/cm (wire grid 3780 V, 1st ring 3580 V, cathode 3780 V) [Yi: 1st ring setting is incorrect, inverted fied between wire grid and field cage ring]
Threshold: 400 ADC below baseline (baselines equalized to 15500)
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Comment: |
85
|
05 Nov 2019 08:32 |
Alex / Sam | Routine | General | Run 900, Edrift 500V/cm | Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar) [Edgar\Yi: correction: OFF (during recirculation tpc pressure about 400 mbar)]
Fields: drift 500 V/cm, extraction 2.8 kV/cm (wire grid 3780 V, 1st ring 2780 V, cathode 7780 V) [Yi: 1st ring setting is incorrect, inverted fied between wire grid and field cage ring]
Threshold: 400 ADC below baseline (baselines equalized to 15500)
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Comment: |
86
|
05 Nov 2019 23:58 |
Yi Wang | Routine | General | Morning data taking, purity check | Run 917
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 918
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 919
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 920
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
|
88
|
06 Nov 2019 11:22 |
Yi Wang | Routine | General | Evening data taking, purity check | Not sure if the trigger threshold is correctly set, there is an error:
Traceback (most recent call last):
File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 71, in <module>
main()
File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 47, in main
(mean, rms) = fit_res
ValueError: too many values to unpack (expected 2)
the trigger threshold for all the runs in this Elog is 2699 ADCc below the baseline (see Elog 89)
Run 943
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 945
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 946
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 948
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 949
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 951
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 952
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 954
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1600 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k |
89
|
06 Nov 2019 12:28 |
Ben Smith | Problem Fixed | General | Evening data taking, purity check | > Not sure if the trigger threshold is correctly set, there is an error:
Sorry, I made a change to the baseline calculation and forgot to update the threshold tool. It's updated now.
The thresholds currently seem to be set to 12801 - this is 2699 below the nominal baseline of 15500. |
90
|
07 Nov 2019 02:07 |
Sam | Routine | General | Run 963 Laser Run with SiPMs at 65V, 12us trace | Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V |
91
|
07 Nov 2019 02:07 |
Yi/Marco/Andrea | Routine | General | morning data taking, purity check | Run 956
Gas pocket: ON (thickness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 957
Gas pocket: ON (thickness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 958
Gas pocket: ON (thickness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 960
Gas pocket: ON (thickness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 961
Gas pocket: ON (thickness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
TPC Pressure: >200 mbarg
SiPM HV: 65 V
Number of events: 10k |
92
|
07 Nov 2019 02:16 |
Sam | Routine | General | Run 964 Laser Run with SiPMs at 65V, 30us trace | Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 30us, 5us pre-trigger
SIPM HV: 65V |
93
|
07 Nov 2019 02:34 |
Sam | Routine | General | Run 966 Laser Run with SiPMs at 65V, 30us trace | Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 30us, 5us pre-trigger
SIPM HV: 65V
MISSING FILE (E.Sanchez) |
94
|
07 Nov 2019 23:48 |
Yi Wang | Routine | General | morning data taking | Run 973
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 222 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 974
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 222 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 975
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 222 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 976
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 222 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 978 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 979 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
After laser run:
Switch Laser OFF
In Chronobox:
DIsable "External Triger Enable [ext_trig_ena]0\[0]
Input "0x3F3F3F7F" in "Enable Channel [ch_enable]
|
101
|
08 Nov 2019 10:07 |
Edgar Sanchez | Routine | General | Monitoring Data Taking | Run 995
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 230 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz
Comment: TPC pressure increased up to 290 mbarg
Run 996
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 225 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz
Comment: TPC pressure increased up to 240 mbarg
Run 997
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz
Run 998
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 10k
Rate ~9 Hz |
104
|
09 Nov 2019 00:13 |
Xiang Xiao | Routine | General | Morning data taking | "ERROR INFO: 10:38:10.924 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."
Emailed to Ben.
Run 999
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 233 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1000
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 235 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1001
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 238 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1002
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 238 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1003 Laser run (manually stopped)
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Comment: Triger rate too low (~8 Hz), manually stopped
Run 1004 Laser run (manually stopped)
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Comment: Triger rate too low (~8 Hz), manually stopped
|
106
|
09 Nov 2019 09:39 |
Sam Hill | Routine | General | Evening data taking | Run 1016
Gas pocket: ON (thinkness 7mm)
Fields: Off
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 201 mbarg
SiPM HV: 65 V
Number of Events: 10K
Run 1017
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 201 mbarg
SiPM HV: 65 V
Number of Events: 10K
Run 1018
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 217 mbarg
SiPM HV: 65 V
Number of Events: 10K
Run 1019
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240
SiPM HV: 65 V
Number of Events: 10K
Run 1020 Laser Run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Comment: Recieved this error message but run seemed to continue ok and data appeared to be written: 19:56:04.839 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run. ╳
Run 1021 Laser Run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
|
107
|
10 Nov 2019 00:06 |
Yi Wang | Routine | General | morning data taking | Run 1022
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 159 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1024
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 186 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1025
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 207 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1026
Gas pocket: ON (thinkness 7mm)
Fields: OFF
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 225 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1028 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k
Run 1029 Laser run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Number of events: 100k |
108
|
10 Nov 2019 08:04 |
Edgar Sanchez | Routine | General | S1 data taking with field | Run 1032
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 800 Hz
Run 1033
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 350 Hz
Run 1034
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 250 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 260 Hz
Run 1035
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 400 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 200 Hz
Run 1036
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 600 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 130 Hz
"ERROR INFO: 17:19:23.176 2019/11/10 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run."
Run 1037
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 800 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 110 Hz
Run 1038
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 80 Hz
Run 1039
Gas pocket: OFF
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 150 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
|
109
|
10 Nov 2019 08:40 |
Edgar Sanchez | Routine | General | S1 data taking without field | Run 1041
Gas pocket: OFF
Fields: OFF
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 360 Hz
Run 1042
Gas pocket: OFF
Fields: OFF
Threshold: 200 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 340 Hz
Run 1043
Gas pocket: OFF
Fields: OFF
Threshold: 250 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 290 Hz
Run 1044
Gas pocket: OFF
Fields: OFF
Threshold: 400 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 240 Hz
Run 1045
Gas pocket: OFF
Fields: OFF
Threshold: 600 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 170 Hz
Run 1046
Gas pocket: OFF
Fields: OFF
Threshold: 800 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 140 Hz
Run 1047
Gas pocket: OFF
Fields: OFF
Threshold: 1000 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 110 Hz
Run 1048
Gas pocket: OFF
Fields: OFF
Threshold: 150 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
Run 1049
Gas pocket: OFF
Fields: OFF
Threshold: 100 ADCc below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 340 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz |
110
|
10 Nov 2019 11:18 |
Edgar Sanchez | Routine | General | Afternoon system monitoring | Run 1050
Gas pocket: ON (unknow)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 245 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1051
Gas pocket: ON (unknow)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 10k
Run 1052
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 245 mbarg
SiPM HV: 65 V
Number of events: 10k
Stopped. Problem with DAQ |
111
|
10 Nov 2019 11:19 |
Edgar Sanchez | Routine | General | S1 + S2 short time window | Run 1054
Gas pocket: ON (unknow)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 100 ADC below baseline
Trace length: 30us total, 5 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240 mbarg
SiPM HV: 65 V
Number of events: 50k
Rate: 400 Hz
|
115
|
11 Nov 2019 01:12 |
Yi / Edgar | Routine | General | Morning data taking | Run 1065
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADCc below baseline
Trace length: 100us total, 80 us post-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1066
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 500 ADCc below baseline
Trace length: 100us total, 80 us post-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 20k
Run 1067
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 200 ADCc below baseline
Trace length: 100us total, 80 us post-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 260 mbarg
SiPM HV: 65 V
Number of events: 20k |
|