Run 1016
Gas pocket: ON (thinkness 7mm)
Fields: Off
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 201 mbarg
SiPM HV: 65 V
Number of Events: 10K
Run 1017
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 2.8 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 201 mbarg
SiPM HV: 65 V
Number of Events: 10K
Run 1018
Gas pocket: ON (thinkness 7mm)
Fields: drift 200 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 217 mbarg
SiPM HV: 65 V
Number of Events: 10K
Run 1019
Gas pocket: ON (thinkness 7mm)
Fields: drift 500 V/cm, extraction 3.78 kV/cm
Threshold: 1000 ADC below baseline
Trace length: 200us total, 100 us pre-trigger
Threshold extend: 5x16 = 80ns
TPC Pressure: 240
SiPM HV: 65 V
Number of Events: 10K
Run 1020 Laser Run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MAX
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
Comment: Recieved this error message but run seemed to continue ok and data appeared to be written: 19:56:04.839 2019/11/09 [feov1725MTI00,ERROR] [feoV1725.cxx:1048:read_trigger_event,ERROR] Error: did not receive a ZMQ bank after 100.000000 ms. Stopping run. ╳
Run 1021 Laser Run
Chronobox: External trigger Enable
Chronobox: Enable Channel [ch_enable]: 0x00000000
Laser Intensity: MIN
Threshold: external clock (laser NIM), laser set to 1kHz
Trace length: 12us, 5us pre-trigger
SIPM HV: 65V
|