HW-Time-Slice Firmware-Filter Software vx2740 Steering Module Reports MVM MVM Vexos MVM-Bug listing MVM TRIUMF Local DS Prototype DS Cryogenic For Shifters BCIT-31 ChronoBox Run Operation DS-DAQ
  CERN DS-Proto0 read-only backup, Page 1 of 8  Not logged in ELOG logo
ID Date Authordown Type Category Subject
  91   07 Nov 2019 02:07 Yi/Marco/AndreaRoutineGeneralmorning data taking, purity check

Run 956

Gas pocket: ON (thickness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 957

Gas pocket: ON (thickness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 958

Gas pocket: ON (thickness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 960

Gas pocket: ON (thickness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 961

Gas pocket: ON (thickness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

  131   14 Nov 2019 04:03 Yi/Edgar/SamRoutineGeneralNoise data taking

Run 1121 

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 50 V (lower than breakdown voltage) 

Number of events: 5k

 

Run 1122 

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 63 V

Number of events: 5k

Without blanket 

 

Run 1125

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 5k

Without blanket 

 

Run 1126

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 67 V

Number of events: 3.5k

trigger rate went down to 1 Hz, run was stopped manually

Without blanket 

 

Run 1127

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 67 V

Number of events: 2k

Without blanket 

 

Run 1128

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 67 V

Number of events: 5k

With blanket 

trigger rate went down to 1 Hz, run was stopped manually

problem with DAQ forced to restart manually 

Run 1129

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 67 V

Number of events: 5k

With blanket 

trigger rate went down to 1 Hz, run was stopped manually

problem with DAQ forced to restart manually 

jsroot_server stops from time to time 

1573739641 14:54:01.099 2019/11/14 [Logger,INFO] Client 'jsroot_server' on buffer 'SYSTEM' removed by cm_periodic_tasks because process pid 26693 does not exist

Run 1130

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 67 V

Number of events: 5k

With blanket 

 

Run 1131

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 5k

With blanket 

Run 1132

Gas pocket: OFF

Fields:  OFF

Threshold: external clock, 50Hz

Trace length: 1ms total, 600 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 5k

With blanket 

 

  71   05 Nov 2019 04:48 Yi WangRoutineGeneralRun 873

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar)

Fields: drift 500V/cm extraction 3.78kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment: try to trigger on S2

  72   05 Nov 2019 04:57 Yi WangRoutineGeneralRun 875

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment: compare to Run 873

  73   05 Nov 2019 05:09 Yi WangRoutineGeneralRun 877

Gas pocket: ON (thinkness unknown, since the TPC pressure is >200 mbar)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Comment: compare to Run 873, 875

  86   05 Nov 2019 23:58 Yi WangRoutineGeneralMorning data taking, purity check

Run 917

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 918

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 919

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 920

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

  88   06 Nov 2019 11:22 Yi WangRoutineGeneralEvening data taking, purity check

Not sure if the trigger threshold is correctly set, there is an error:

Traceback (most recent call last):
  File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 71, in <module>
    main()
  File "/home/dsproto/online/dsproto_analyzer/threshold_tool.py", line 47, in main
    (mean, rms) = fit_res
ValueError: too many values to unpack (expected 2)

the trigger threshold for all the runs in this Elog is 2699 ADCc below the baseline (see Elog 89)
 

Run 943

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 945

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 946

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 948

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 949

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 951

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 952

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 954

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1600 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

TPC Pressure: >200 mbarg

SiPM HV: 65 V

Number of events: 10k

  94   07 Nov 2019 23:48 Yi WangRoutineGeneralmorning data taking

Run 973

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 974

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 975

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 976

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  222 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 978 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 979 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

After laser run:

Switch Laser OFF

In Chronobox:

DIsable "External Triger Enable [ext_trig_ena]0\[0]

Input "0x3F3F3F7F" in "Enable Channel [ch_enable] 

 

  107   10 Nov 2019 00:06 Yi WangRoutineGeneralmorning data taking

Run 1022

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  159 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1024

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  186 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1025

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  207 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1026

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADC below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  225 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1028 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 1029 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock (laser NIM), laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

  113   11 Nov 2019 00:37 Yi WangProblemMIDAScannot connect to rootana

The connection to rootana is lost.

No data taking can be started.

  118   11 Nov 2019 23:57 Yi WangProblem FixedTriggerID 117 problem is fixed

The problem reported in ID 117 is somehow fixed this morning.

  119   12 Nov 2019 00:16 Yi WangProblemDigitizerV1725 error, runs keep crashing

09:15:33.955 2019/11/12 [feov1725MTI00,ERROR] [feoV1725.cxx:663:link_thread,ERROR] Exiting thread 3 with error

  121   12 Nov 2019 00:24 Yi WangRoutineGeneralMorning data taking

Run 1084

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1085

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1086

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  240 mbarg

SiPM HV: 65 V

Number of events: 6k

Run crashed: 09:31:55.239 2019/11/12 [feov1725MTI00,ERROR] [feoV1725.cxx:663:link_thread,ERROR] Exiting thread 0 with error

  125   13 Nov 2019 00:08 Yi WangRoutineGeneralMorning data taking

Run 1099

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  250 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1100

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  250 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1101

Gas pocket: ON (thinkness 7mm)

Fields: drift 500 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  250 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1102

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  250 mbarg

SiPM HV: 65 V

Number of events: 10k

 

Run 1103

Gas pocket: ON (thinkness 7mm)

Fields: OFF

Threshold: 1000 ADCc below baseline

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  250 mbarg

SiPM HV: 65 V

Number of events: 10k

  134   15 Nov 2019 00:19 Yi WangRoutineGeneralLaser run

Run 1166 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock, laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

Trigger rate went to 0, run was stopped manually.

 

Run 1169 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MAX

Threshold: external clock, laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

Run 1170 Laser run

Chronobox: External trigger Enable

Chronobox: Enable Channel [ch_enable]: 0x00000000

Laser Intensity: MIN

Threshold: external clock, laser set to 1kHz

Trace length: 12us, 5us pre-trigger

SIPM HV: 65V

Number of events: 100k

 

  136   16 Nov 2019 01:57 Yi WangRoutineGeneralDual phase data taking proposed by Marco

Run 1186

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 100 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 10k

V1725 offline, run stopped manually

 

Run 1188

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 100 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 70k

Trigger rate became very low, run stopped manually

 

Run 1189

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1190

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  10

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1191

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  15

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1192

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  5

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1193

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  10

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1194

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 2

Trace length: 30us total, 5 us pre-trigger

Threshold extend:  15

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1195

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 2

Trace length: 200us total, 100 us pre-trigger

Threshold extend:  5

TPC Pressure:  26 mbarg

SiPM HV: 65 V

Number of events: 10k

  137   17 Nov 2019 00:48 Yi WangRoutineGeneraldual phase data taking

Run 1197

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1198

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 2.8 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1200

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1201

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

Data taking has to be stopped because of the noises.

 

Attachment 1: Screenshot_from_2019-11-17_11-02-26.png
Screenshot_from_2019-11-17_11-02-26.png
  138   17 Nov 2019 02:23 Yi WangRoutineGeneraldual phase data taking

Noise problem seems like fixed after remaking the Al foil cover.

Run 1208

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1209

Gas pocket: ON (thinkness 9mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Coincidence: 3

Trace length: 100us total, 50 us pre-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  30 mbarg

SiPM HV: 65 V

Number of events: 20k

At the end of run 1209, a lot of noise show up again.

  114   11 Nov 2019 00:47 Yi / EdgarProblem FixedDigitizerProblem fixed by restarting the VME crate
  115   11 Nov 2019 01:12 Yi / EdgarRoutineGeneralMorning data taking

Run 1065

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 1000 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1066

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 500 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

 

Run 1067

Gas pocket: ON (thinkness 7mm)

Fields: drift 200 V/cm, extraction 3.78 kV/cm

Threshold: 200 ADCc below baseline

Trace length: 100us total, 80 us post-trigger

Threshold extend:  5x16 = 80ns

TPC Pressure:  260 mbarg

SiPM HV: 65 V

Number of events: 20k

ELOG V3.1.4-cb3afcd8