Message ID: 64     Entry time: 03 Nov 2019 09:17
Author: Alex Kish 
Type: Routine 
Category: General 
Subject: run 836, zero drift field 

Gas pocket: ON

Fields: drift field OFF (1st ring 0 V, cathode 0 V), extraction field OFF (wire gate 0 V)

Threshold: 200 ADCc below baseline

Threshold extend:  5x16 = 80ns

TPC Pressure: 244 mbarg

SiPM HV: 50 V (noise run)

Comment: noise conditions good. Thresholds are too high, rate below 1 event/sec, only few events saved. Postpone to next day.