U/I QPump Monitoring

Purpose

Monitors the global Bias voltage and Bias current.

Description

Two 10 bit channels of the uC internal ADCs provide readback of the charge pump voltage and current.

Voltage monitoring

Charge Pump Voltage monitoring schematic
vQPump.jpg
Table of values associated with the Voltage QPump measurements
R1/%         R2/%       #of bits       VREF    1/Gain   ADCgain    V/DAC
1M/0.1%   10K/0.1%     10 bits(1024)  2.43V     101       2        0.1198

Current monitoring

Charge Pump Current monitoring schematic
iQPump.jpg
Table of values associated with the currrent QPump measurements
Isource    R1    A/V    #of bits          VREF       A/DAC
IBias     100R    1.    10 bits (1024)    2.43V     0.00002373

DataSheet

Current monitor HV7800


Generated on 1 Feb 2022 for FGDC8051-FEB64 by  doxygen 1.6.1