Monitors each channel's voltage/current and compare them with the global Voltage and current Bias. The sum of the channels' currents should match the Qpump current value. When each channel is turned on, its voltage should also be approximately equal to the VBias. Any discrepancy indicates a potential current leak or disconnection.
Presently the procedure described above is not yet implemented
R1/% R2/% R3/% #of bits VREF (VREF/1024) Attenuation 100R/0.1% 976k/0.1% 24.9k/0.1% 10 bits(1024) 2.43V 0.002373 0.0248776 R1 A/V #of bits VREF A/DAC IBMon 100R 1. 16 bits (65536) 1.25V 5.96E-9