  8   28 Jan 2019 01:44 ThomsRoutineSoftwareAdded web display of V1725 waveformsI added javascript webdisplay of the V1725
waveforms to the Darkside setup.  You can
see the waveform display here:
  
  7   21 Jan 2019 13:05 Thomas, PierreRoutineDigitizerTests of V1725 baseline and triggeringA little progress on a couple fronts:

1) We got the baseline DAC setting working
  
  6   17 Jan 2019 15:47 ThomasRoutineHardwareAdded raided HD to ds-proto-daqAdded pair of 10TB hard-drives in raid-1 to
ds-proto-daq.  MIDAS data files will get
written to this raid volume.
  
  5   11 Jan 2019 11:07 ThomasRoutineDigitizerData corruption for ADC channelsI modified the analyzer so that it shows data
from all four modules.
 dsproto_corrupt.gif 
  4   09 Jan 2019 12:22 ThomasRoutineGeneralV1725 LVDS outputsPierre figured out that NIM crate not working.
 We now see LVDS outputs from the individual
channels firing. 
  
  3   13 Dec 2018 11:02 ThomasRoutineHardwareTesting V1725 digitizersCouple weeks of work documented in one elog...

V1725 serial numbers: 455, 392, 460, 462,
  
  2   11 Dec 2018 15:20 ThomasConfigurationSoftwareSetup elog for ds-proto-daq
1) Install and tweak elog:
  
  1   11 Dec 2018 15:17 ThomasConfigurationSoftwareSetup elog for ds-proto-daq
1) Install and tweak elog:
  